The unit is equipped to store more than 5000 test results and export to an optional printer or to a LIMS
| Standards | ASTM D92, ISO 2592, IP 36, GB/T 3536 |
| Measuring Range | 79–370°C(175–698°F) |
| Resolution | 0.1°C |
| Repeatability | ≤6°C |
| Reproducibility | ≤8°C |
| Safety Features | Automatically shuts down, alarm, and cooling, in case: • Flash Point Mode: EFP + 20°C is reached • Flash & Fire Mode: EFP + 50°C is reached • 400°C is reached |
| Gas Supply | LP gas, butane gas, pipeline gas or natural gas (pressure < 10kPa) |
| Cooling | Air cooling |
| Ignition Source | Gas ignition with automatic lighting |
| Flash Point Detection | Ionization Ring |
| Temperature Sensor | PT100 probe |
| User interface | 7" color touch screen Languages:Chinese, English, Russian |
| Barometric correction | Flash point is corrected to barometric pressure |
| Data Storage | No less than 5000 results |
| I/O Port | RS232, RJ-45 |
| Data Export | LIMS or Printer |
| Machine Power | 900W |
| Heater Power | 800W |
| Operating Conditions | 5°C–35°C 20–90%RH at 35°C, Non-condensing No air flow |
| Power Supply | AC 220V±10% 50/60Hz |
| Dimensions | 530mm(W) × 233mm(D) × 427mm(H) |
| Weight | 19KG |