BM-8850 Gas Chromatograph


1. It adopts an advanced 100/1000 Mbps Ethernet communication interface with a built-in

IP protocol stack, allowing the instrument to achieve long-distance data transmission through

the enterprise's internal local area network and the internet; facilitating laboratory setup, simplifying

 laboratory configuration, and convenient management of analytical data. 


2. The instrument is equipped with a 7-inch color LCD touch screen, which supports hot-swapping

and can be used as a handheld controller. 


3. The instrument adopts a multi-processor parallel working mode, making the instrument

more stable and reliable; it can be optionally equipped with various high-performance

detectors, such as FID, TCD, ECD, FPD, and NPD, with a maximum of three detectors 

installed simultaneously.


 

1. It utilizes advanced 100/1000 Gigabit Ethernet communication interfaces and a built-in IP protocol stack, allowing the instrument to achieve long-distance data transmission through the enterprise's internal local area network and the internet; this facilitates laboratory setup, simplifies laboratory configuration, and simplifies the management of analytical data.

2. The instrument is equipped with a 7-inch color LCD touch screen, which supports hot-swapping and can be used as a handheld controller.

3. The instrument adopts a multi-processor parallel working mode, making it more stable and reliable; it can be optionally equipped with various high-performance detectors, such as FID, TCD, ECD, FPD, and NPD, with up to three detectors installed simultaneously. The FID uses a logarithmic amplifier and has an automatic ignition function.

4. The brand-new microcomputer temperature control system offers high temperature control accuracy, superior reliability, and anti-interference performance; it has eight completely independent temperature control outputs, enabling 50-step programmed temperature control; it features an automatic column oven door opening system, improving near-room temperature control capabilities and achieving faster heating/cooling speeds. Each heating control unit has over-temperature protection.

5. The instrument's dedicated workstation can simultaneously support multiple chromatographs, enabling data processing and reverse instrument control, simplifying document management.

6. The chromatograph has a built-in low-noise, high-resolution 24-bit AD circuit and features baseline storage and baseline subtraction functions.

7. The chromatograph microcomputer system, with complete independent intellectual property rights, uses the standard MODBUS/TCP protocol and can interface with DCS systems.

8. Full EPC electronic flow control system, with gas protection and electronic flow control from 0 to 1500 ml/min.
● Operation Display: 7-inch color LCD touch screen, can be used as a handheld controller
● Temperature Control Zones: 8 channels; Temperature Control Range: Room temperature +4℃~460℃, Accuracy: ±0.1℃; Programmed temperature control: 50 steps.
● External Events: 8 channels; Auxiliary control outputs: 2 channels
● Injector Types: Packed column injection, capillary injection, six-port valve gas injection, Automatic sampler (optional)
● Number of Detectors: 3 (maximum); FID, TCD, ECD, FPD, and NPD (selectable)
● Injection Start: Manual or automatic (selectable)
● Communication Interface: Ethernet: IEEE802.3
● EPC (Electronic Pressure and Flow Control) system